Features:
- Height × width × depth: 185 × 605 × 424 mm (7.3 × 23.8 × 16.7 inches)
- Weight: 15.1 kg (33.2 lb)
- Line voltage (selectable): 80–132 or 170–264 V AC
- Line frequency: 47–66 Hz
- Power consumption: 71 VA typical
- Operating temperature: 0–40 °C (32–104 °F)
- Non-operating temperature: -40–65 °C (-4–149 °F)
Specifications:
Wavelength range | 190–820nm | – |
Slit width | 2 nm | – |
Stray light | <0.05 % | at 220 and 340 nm, Hoya O56 |
Wavelength accuracy | <± 2 nm | NIST 2034 |
Wavelength reproducibility | <±0.05 nm | NIST 2034, ten consecutive measurements |
Photometric accuracy | ±0.005 A | at 1 A, 440 nm, NIST 930e |
Photometric noise | <0.0002 A | at 0 A, 500 nm, sixty 0.5-second scans, rms |
Photometric stability | <0.002 A/hr | (at 0 A, constant ambient temp., after 1-hour warm up):measured over 1 hour, every 5 seconds |
Baseline flatness | <0.001 A | 0.5 second blank, 0.5 second scan, rms |
Typical scan time | 0.5 s | full range |
Shortest scan time | 0.1 s | full range |
Time until next scan | 0.6 s | no variance/0.1 second |