Type |
Specification
|
Pressure |
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5800psi) |
GLP features |
Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications |
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range |
0.1 – 900 µl in 0.1 µl increments (recommended 1 µl increments) Up to 1800 µl with multiple draw (hardware modification required) Up to 5000 µl with multiple draw (hardware modification required) |
Replicate injections |
1 – 99 from one vial |
Precision |
Typically < 0.5% RSD of peak areas from 5 – 2000 µl, Typically < 1% RSD of peak areas from 2000 – 5000 µl, Typically < 3% RSD of peak areas from 1 – 5 µl |
Minimum sample volume |
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial |
Sample viscosity range |
0.2 – 50 cp |
Sample capacity |
100 × 2-ml vials in 1 tray 15 × 6-ml vials in ½ tray (Agilent vials only) |
Injection cycle time |
Typically 50 s, depending on draw speed and injection volume |
This refurbished HP / Agilent 1100 Series G2260A ALS Prep Autosampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Multiple available.