This Agilent 1200 Series 2258 A Dual Loop Autosampler is designed for flexible sample handling for both analytical (10 mL/min flow rate) to preparative scale purification (100 mL/min flow rate). It is in good condition and has been tested at SpectraLab Scientific.
A 90-day warranty is included.
We are professionals. We encourage you to contact us to speak with one of our expert and professional team members. Thanks for looking and we look forward to hearing from you!
Type | Specification |
GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications | Controller-area network (CAN). RS232C, APG-remote standard, optional: four external contact closures and BCD vial number output optional: LAN communication card |
Instrument control | Agilent ChemStation software Agilent HT Purification SW Agilent G1323B Control Module |
Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection modes | Partial loop filling Complete loop filling |
Injection range | 0.1 – 5000 µl in 0.1 µl increments Recommended injection range 10 – 5000 µl |
Area precision Complete Loop Fill | Typically < 0.5% RSD,(50 µl loop, overfill factor 3 or greater) |
Area precision Partial Loop Fill | Typically < 1.0% RSD,(5000 µl loop, 1000µl injection volume, draw/eject speed 20 ml/min) |
Carry-over | < 0.1% with rinse factor > 3 (see test method listing) |
Area precision Complete Loop Fill | Typically < 0.5% RSD,(50 µl loop, overfill factor 3 or greater) |
Area precision Partial Loop Fill | Typically < 1.0% RSD,(5000 µl loop, 1000 µl injection volume, draw/eject speed 20 ml/min) |
Draw/eject speed | max. 50 ml/min (selectable in 1 µl/min steps) |
Loop size | max. loop size 5 ml (selectable range: 5 µl – 50000 µl) |
Sample viscosity range | 0.2 – 5 cp |
Injection volume | Selectable range: 0 – 50000 µl in 0.1 µl steps Recommended range: 10 µl – 5000 µl in 1 µl steps |
Temperature controlled version with G1330B |
Temperature range of sample compartment: 4 – 40 deg C |
Sample capacity | 2 × well-plates (MTP) + additional 10 × 2 ml vials |
Supported sample Plates |
Any combination of 2 of the following plates: 96 well plate (shallow/deep/conical) 54 x 2 ml vial plate 15 x 6 ml vial plate 27 x Eppendorf plate for Eppendorf tubes of 0.5 ml,1.5 ml or 2.0ml Maximum plate height with plate tray G2258-60001: 50 mm |