The Accela Autosampler holds up to 200 standard vials, three 96-well plates, or three 384-well plates; contains a built-in column oven (5 to 95°C); provides tray/sample temperature control (0 to 60°C). It\’s also capable of performing automated sample preparation routines.
S/N: 82930
Many in Stock
Specifications:
Type | Specifications |
---|---|
Sample capacity | 200 vials in five removable trays 288 samples in three 96-well microplates 1152 samples in three 384-well microplates |
Standard vial | 1.8 mL volume |
Injection precision | < 1.0% RSD at 5 μL and up |
Partial loop injection volume | 10 μL with standard 25 μL sample loop and 250 μL syringe Up to 1000 μL injection volume with optional loops and syringes |
Fixed loop (full loop) injection volume | 25 μL with standard loop and 250 μL syringe Up to 1000 μL injection volume with optional loops and syringes |
Sample carryover | < 0.1% 6 mL flush volumes, user-specified for internal and external needle wash cycles |
Minimum sample volume | 1 μL can be injected from 5 μL |
Needle height control | Programmable in 0.1 mm increments Active vial bottom search selectable on/off |
Column oven | Programmable in 1 °C increments from 5 to 95 °C |
Tray temperature control | Programmable in 1 °C increments from 0 to 60 °C Typical vial temperature from 4 to 50 °C |
Reagent vials | Four 16 mL vials can be specified for flush/wash or dilution functions |
Remote inputs | Pump ready, inject hold |
Remote outputs | Pump stop, autosampler ready, inject, gradient start, four timed events |
Ambient environment | 10 to 40 °C, 5 to 95% relative humidity, non-condensing |
Dimensions | 37x36x50cm |