The TOC analyzers adopts the 680°C combustion catalytic oxidation method, which was developed by Shimadzu and is now used worldwide. While providing an ultra wide range of 4 μg/L to 30,000 mg/L, these analyzers boast a detection limit of 4 μg/L through coordination with NDIR. This is the highest level of detection sensitivity available with the combustion catalytic oxidation method. In addition, the combustion catalytic oxidation method makes it possible to efficiently oxidize not only easily-decomposed, low-molecular-weight organic compounds, but also hard-to-decompose insoluble and macromolecular organic compounds.
Measurement range | 4μg/L to 4,000mg/L | |
Automatic dilution | None | |
NPOC measurement time (No sampler) |
10.5 minutes, sample fractionated into vials -> acid added -> sample set -> conditions set -> measurement | |
IC pretreatment | Sparging only, automated treatment outside the analyzer | |
Rinsing water | None, replacement rinsing by sample (rinsing with sample ) only | |
COD/BOD conversion | Function not available | |
Total nitrogen measurement | Not available | |
Built-in LCD | Black and white | |
Accuracy control function | None | |
ER/ES | None | |
Number and type of data saved | Measurement results: 200 sample average values can be saved | |
Output of saved data | Not available | |
NDIR light intensity adjustment | Required | |
Carrier gas problem detection | Function not available | |
Main unit dimensions, weight | W480×D500×H480mm,Approx.46kg | |
Autosampler sample vials | ASI-5000A:5mL×78,42mL×16 (as well as STD 8), special vials used | |
Simple autosampler | None set | |
Halogen scrubber for high-salt samples (B-type halogen scrubber) | No option available |