Agilent 1100 /MSD (G1946D)
Mass Range
- 2-3000 m/z
Mass Accuracy
- +- 0.13 u within the calibrated mass range in scan mode
Mass Axis Stability
- +- 0.13 u over 8 hours
Scan Speed
- 2500 u/s standard mode
- 5250 u/s fast scan mode
SIM Sensitivity
- ESI at 400 ul/min
- APCI at 1000 ul/min
- Selected ion monitoring of m/z 609.3
- Positive ionization
- Signal To Noise Ratio
- Quantity: 1 pg reserpine 100:1 RMS (20:1 peak-to-peak)
Scan Sensitivity
- ESI at 400 ul/min
- APCI at 1000 ul/min
- Scan range m/z 100-650
- Scan speed 2500 u/s
- Extracted ion at m/z 609.3
- Positive ionization
- Signal To Noise Ratio
- Quantity: 50 pg reserpine 100:1 RMS (20:1 peak-to-peak)
Multiple Signal Acquisition
- Ability to cycle through four different acquisitive on modes on a scan-by-scan basis within a single run
The fully tested and refurbished G1946D comes with ESI, Rough pump and suitable software and PC
And also Front End Agilent 1200(G1312B, G1316A, G1379B with solvent tray) or 1100 HPLC with CTC HTC Pal can be included as options.
We are professional to the stuffs. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!