Type |
Specification |
Injection modes |
Partial loop filling Complete loop filling |
Injection range |
0.1 – 5000 µL in 0.1 µL increments. Recommended injection range 10 – 5000 µL |
Loop size |
Max. loop size 5 ml (selectable range: 5 µL – 50000 µL) |
Area precision Complete loop fill |
Typically < 0.5% RSD (50 µL loop, overfill factor 3 or greater) |
Area precision Partial loop fill |
Typically < 1.0% RSD (5000 µL loop, 1000 µL injection volume, draw/eject speed 20 mL/min) |
Sample viscosity range |
0.2 – 5 cp |
Sample capacity |
2 well-plates (96 and 384) plus 10 additional 2-mL vials 108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2-mL vials 30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2-mL vials 54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 x 27 Eppendorf tube plate Vial septa and closing mats must be pre-slit. |
Carryover |
< 0.1% with rinse factor > 3 (see test method listing in manual) |
This Agilent 1200 G1329B ALS SL Autosampler is in good shape.
A 90 day warranty is included.
We encourage you to contact us to speak with one of our expert and professional team members. Thanks for looking and we look forward to hearing from you!