The Dionex PDA-100 Photodiode Array Detector is an optical detector capable of measuring the absorbance spectrum from 190 nm to 800 nm. A deuterium lamp optimizes the UV range (190 nm to 380 nm) and a tungsten lamp optimizes the visible range (380 nm to 800 nm).
When using the PDA-100, up to five single wavelengths (2D chromatograms) can be collected without being required to collect 3D data. Collecting individual wavelengths instead of the spectra offers two advantages: it eliminates the need to perform extractions for runs that do not require spectral data and it conserves disk space.
This refurbished Dionex PDA-100 Photodiode Array Detector is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We are seasoned professionals in this area and we would be happy to have you speak with one of our experts here at SpectraLab.
Thanks for looking and we look forward to hearing from you!
Specification | |
Optical System | Single-beam, reverse-optics design with concave holographic Grating |
Light Sources | Deuterium lamp (30 W) for ultraviolet spectrum analysisTungsten lamp (15 W) for visible spectrum analysis |
Photodiode Array | 1024 elements |
Wavelength Range | 190 nm to 800 nm |
Pixel Resolution | 0.7 nm |
Wavelength Accuracy | ± 1 nm |
Noise | ± 10 µAU with 2-second rise time, 254 nm, 4 nm bandwidth, and flowing deionized water± 15 µAU with 2-second rise time, 520 nm, 10 nm bandwidth, and flowing deionized water |
Drift | < 500 µAU/hour (after warm-up) |
Linearity | > 2 AU |
Analog Outputs | Four built-in analog outputs |
Voltage Out | 1 V output |
Analog Output Range | 0.0005, 0.001, 0.01, 0.1, 0.5, 1.0, 2.0, or 3.0 AU full-scale |
Control and Data Evaluation | Provided by PeakNet 6 or Chromeleon software; communication with the detector is via the DX-LAN interface |
Cell Body | PEEK or stainless steel |
Volume and Optical Path Length | PEEK cell: 13 µL volume; 10 mm (0.394 in) path lengthStainless steel cell: 13 µL volume; 10 mm (0.394 in) path length |
Maximum Operating Pressure | PEEK cell: 2 MPa (300 psi)Stainless steel cell: 3 MPa (500 psi) |
Chemical Compatibility | -Do not use bases stronger than 0.1 M; they will etch the silica windows of the flow cells.-Do not use normal phase or chlorinated solvents with the PEEK flow cell; they will damage the cell.
-Do not use high pH salts with the stainless-steel flow cell; they will cause corrosion, contamination, and metal leaching. |