Specifications
Type | Specifications |
---|---|
Method of Detection | Deflection |
Light source | Tungsten lamp |
Refractive index range | 1.0 to 1.75 |
Baseline noise | 2.5 x 10-9 RIU or less (water filled in, time constant 3.0 s) |
Temperature control | Off, 30 to 50 |
Time constants | 0.1, 0.25, 0.5, 1, 1.5, 2, 3, 6 s |
External outputs | Warning when measurement range is exceeded, Auto zero, Alarm, Recorder 10 mV/FS, Integrator 1 V/FS |
External inputs | Auto zero, Marker, Purge |
Baseline adjustment range | 0 to 50% /FS |
Auto zero | Full auto zero (optics + calculation) |
Recorder range | 0.25 to 512 x 10-6 RIU/FS |
Cell volume | 8mL |
Max. flow rate | 10 mL/min |
Max. pressure in cell | 7.1 psi (0.049 MPa) |
Dimensions (W x D x H) | 13.5 x 15.75 x 6.0 in |
Weight | Approx.29 lbs (13kg) |