The HP 8452A Diode-Array Spectrophotometer is a single-beam, microprocessor-controlled spectrophotometer. Visible/UV – range of 190 to 820 nm with 2 nm resolution. It is used when speed of measurement is essential. It is faster, more sensitive and some believe it has more precision than a conventional spectrometer due to the photo diode-array detection system.
With its diode array technology, the spectrophotometer is much faster then conventional instruments; with more precision, more sensitivity and more reproducible results.
Features:
- Wavelength Range: 190 to 820 nm
- Wavelength Accuracy: < ± 2 nm
- Wavelength Reproducibility: ± 0.05 nm (typical with constant conditions)
- Spectral Bandwidth: 2 nm
- Stray Light: at 200 nm = < 0.05% , at 340 nm = < 0.05%
- Photometric Range: 0.0022 – 3.3 Å; 0.5 second measurements at 350 nm
- Photometric Accuracy: ±0.005 Å; at 1 Å, 440 nm, NIST 930e
- Photometric Noise: <0.0002 Å RMS; at 0 Å, 500 nm, sixty 0.5-second scans
- Photometric Stability: <0.001 Å/hr; (at 0 Å, constant ambient conditions, after 1 hour warm up): measured over 60 second period, every 5 seconds
- Baseline Flatness: <0.001 Å RMS; 0.5 second blank followed by 0.5 second scan
- Shortest Scan Time: 0.1 seconds; full range
- Rapid Scan Mode: 10 scans per second; 70 nm range
- Power Requirements: Line Voltage: 80-132 or 170-264 V; AC Line Frequency: 47-66Hz
- Dimensions (L x W x H): 23.8 in x 16.7 in x 7.3 in
- Weight: 33.2 lbs.