This refurbished HP / Agilent 1100 Series G1377A μ-WPS AutoSampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Type | Specification |
GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications | Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range | 0.01 – 8 µl in 0.01 µl increments with the small loop capillary 0.01 – 40 µl in 0.01 µl increments with the extended loop capillary |
Precision | Typically < 0.5 % RSD of peak areas from 5 – 40 µl, Typically < 1 % RSD from 1 – 5 µl Typically < 3 % RSD from 0.2 – 1 µl |
Sample viscosity range | 0.2 – 5 cp |
Sample capacity | 2 × well-plates (MTP) + 10 × 2 ml vials 100 x 2 ml in one tray 40 x 2 ml in half tray |
Injection cycle time | Typically < 30 s using following standard conditions: Default draw speed: 4 µl/min Default eject speed: 10 µl/min Injection volume: 0.1 µl |
Carry-over | Typically < 0.05 % using the following conditions: Column: 150 x 0.5 mm Hypersil ODS, 3 µm Mobile phase: Water/Acetonitrile = 85/15 Column Flow rate: 13 µl/min Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |