Type |
Specification |
Sample capacity |
100 x 2 ml vials in 1 tray. Microvials (100 or 300 µl) with sleeves (reduced cooling performance with microvials) |
Settable injection volume |
0.01 to 8 µl with small loop capillary 0.01 to 40 µl with extended loop capillary |
Precision |
Typically < 0.5 % RSD from 5 to 40 µl Typically < 1 % RSD from 1 to 5 µl Typically < 3 % RSD from 0.2 to 1 µl |
Minimum sample volume |
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvials |
Carryover |
Typically < 0.1 % without automated needle wash injection volume |
Sample viscosity range |
0.2 – 5 cp |
Recommended pH-range |
1.0 – 8.5 solvent with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries |
Material in contact with solvent |
Stainless steel, sapphire, PTFE, PEEK, fused silica, Vespel |
GLP features |
Early maintenance feedback (EMF), electronic records of maintenance and errors |
Communications |
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
Safety features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Housing |
All material recyclable |
This refurbished HP / Agilent 1100 Series G1389A MicroALS is in great condition. It has been tested at SpectraLab Scientific.
A 90 days warranty is included.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
Thanks for looking and we look forward to hearing from you!
Multiple available