This Perkin Elmer Vacuum Degasser is sold as-is.
Specification | Description |
Channels | 1-5 independent |
Degassing Process
|
Gas permeation through a fluoropolymer membrane |
Maximum Recommended Flow Rate | 3.0 mL/min. |
Pressure Drop | 1.37 mm Hg/mL/min.
|
Degassing Capacity
|
~25% dissolved gases remaining in 60:40 MeOH/Water mixture at
1 mL/min. |
Dead Volume | ~480 microliters per channel for standard channel
|
Materials contacting solvents | PEEK, Glass-filled PTFE, Teflon AF |
Power: | |
Power requirement if using supplied AC
Adapter
|
100 to 240 VAC (±10%), 1A, 50 to 60 Hz (±3 Hz)
|
Power Requirement if not using supplied AC
Adapter
|
15 to 24 VDC at 0.85 A maximum (0.5 A typical)
|
Wall Sockets | 4 supplied with AC Adapter, interchangeable: North
America/Japan, U.K., Continental Europe, Australia
|
Installation Over-Voltage Category | II |
Validation Output: | |
Signal | 5 mVDC / 1 mm Hg absolute from 20 to 800 mm Hg
(0.100 VDC at 20 mm Hg; 4.000 VDC at 800 mm Hg)
|
Accuracy | ±1.0% of reading ±0.010 VDC from 20 to 800 mm Hg
|
Operating Conditions:
|
|
Ambient Temperature | 10 to 35 ºC
|
Ambient Relative Humidity (RH) | 20 to 80 % RH (without condensation)
|
Altitude | 0 to 2000 Meters
|
Indoor vs. Outdoor Use | Indoor |
Pollution Degree | 2 |
Storage Conditions:
|
|
Ambient temperature | -20 to +60 ºC
|
Ambient Relative Humidity | 20 to 80% RH (without condensation)
|
Altitude | 0 to 12000 M |