The PerkinElmer Clarus 580 GC is a fully automated gas chromatograph. The system offers the capabilities required for laboratories performing large numbers of routine analyses as well as those involved in research and development. All instrument functions are set up and monitored through a touch screen. The intuitive graphical user interface includes a real-time chromatogram display and eight-language support.
The Clarus 580 GC oven provides easy access to columns. The oven gives excellent temperature control and fast cool-down times for high productivity. All temperature and time functions are microprocessor controlled and are shown on the touch-screen display. Software selectable coolant time-out and coolant cut-in temperatures ensure economical subambient operation.
Specifications:
PerkinElmer Clarus 580 GC | |
---|---|
Type | Specifications |
Volume | 10,600 cm3 |
Temperature range | 10 °C above ambient to 450 °C or -99 °C to 450 °C with subambient accessory |
Column overheat protect | User settable up to 450 °C |
Temperature programmer | 3 ramp, 4 plateaus |
Initial oven temperature | Minimum Range: -99 °C to 450 °C Increment: 1 °C |
Initial time | 0 to 999 min Increment: 0.1 min |
Rate | 0.1 to 45 °C/min Increment: 0.1 °C |
Plateau time | 0 to 999 min Increment: 0.1 min |
Final oven temperature | -99 °C to 450 °C Increment: 1 °C |
Cool-down times | 250 °C to 50 °C: 4.8 min 200 °C to 50 °C: 3.8 min 50 °C to 0 °C*: 2.6 min 50 °C to -30 °C*: 3.4 min *with liquid CO2 |