Specification |
|
Light source |
Deuterium lamp, W lamp |
Wavelength range |
190 ~ 900 nm |
Spectral bandwidth |
8 nm |
Wavelength accuracy |
± 1 nm |
Wavelength reproducibility |
± 0.1 nm |
Cell path length |
10nm |
Cell volume |
8 µL |
Cell pressure tolerance |
12MPa |
Dual wavelength mode |
Measurement wavelength: Selectable, two wavelengths in 190- 370 nm or 371-600 nm |
Response |
Selectable in 9 steps corresponding to time constant 0.05, 0.1, 0.5, 1.0, 1.5, 3.0, 6.0, 8.0, 10.0 seconds |
Range |
Can be set between 0.0001 and 2.56 AUFS in 0.0001 AUFS steps |
Zero adjustment |
Auto zero function, baseline shift function |
Wavelength scanning function |
Performs wavelength scanning with flow stopped. Data stored in three files, one of which stores background. Background is subtracted from sample scans. File data is not saved up (lost when power is turned OFF). |
Time Program |
Available in detector or by system controller |
Set items |
Wavelength (including dual wavelengths), auto zero, range, marker, response, wavelength scanning, event, polarity, lamp ON/OFF, loop |
This refurbished Shimadzu SPD-10AVvp UV/vis Detector is in greatcondition. It has been tested at SpectraLab Scientific.
A 90-day warranty is included.
We have special expertise on analytical equipment. We encourageyou to contact us to speak with one of our expert and professional teammembers.
Thanks for looking and we look forward to hearing from you!