This refurbished Shimadzu TOC 5000A Total Organic Carbon Analyzer includes the ASI-5000A autosampler. This system will come as-is.
Analyte | TC, IC, TOC (TC-IC), NPOC (non-purgeable organic carbon), POC (purgeable organic carbon: option)Method Combustion/non-dispersive infrared gas analysis method) |
Combustion temperature | 680° C |
Measuring range | 4ppb to 4000ppm (to 5000ppm for IC) (POC: Max 400ppm) |
Average analysis time | 2 to 3 minutes for both TC and IC (4.5 minutes for TC and 3.5 min. for IC at the longest) |
Repeatability | Standard deviation is less than 1% of full scale for the range less than 2000ppm to 4000ppm (from 2500ppm to 5000ppm for IC) |
Sample introduction | Automatic injection via microliter syringe |
Sample injection volume | 500 to 2000µL with a 2500 µL syringe, 4 to 250 µL with a 250 µL syringe |
Pretreatment for IC | Automatic sparging function |
Carrier gas | Cylinder of ultra-high purity air or oxygen; about 6kgf/cm2 (7kgf/cm2 max.) in supply pressure; 150 mL/min. in flow rate (300mL/min. including sparging) |
Data processing | |
Detector signal | Linearization, baseline correction, peak detection, and area calculation |
Data | Calculation of concentration using one-point to four-point calibration curve; automatic selection of best calibration curve; store up to 18 calibration curves; automatic exclusion of anomalous value(s) and automatic re-analysis/re-calculation; calculation of standard deviation and coefficient of variation, automatic re-analysis with different measurement range and/or different sample volume when response goes off scale. |
Instrument control | Aspiration, measurement, and injection sample; TC furnace temperature; ON/OFF of carrier gas and sparge gas. |
Other features | Automatic selection of the optimum set of operation conditions; maintenance/self-diagnostics, warning of abnormality; timer for restarting from the running mode. |
Display | White liquid crystal display with back-light, 640X400 dots, 88 characters x 22 lines. |
Printer | Thermal, 240-dot graphic printer, 40 characters/line, 110 mm in chart width. External dot matrix: option. |
Ambient temperature requirements | 5° to 35° C |
Power requirements | 100, 110, 115, 120, 127, 220, 230, or 240V as ordered, ±10%, 50/60Hz, about 300 VA power consumption. (800-1200VA.) |