Specification
Heated Devices
- Five heated zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored: Two
Split/Splitless Capillary
- Range to 400°C in 1°C increments
- Back-pressure design permits independent adjustment of split flow rate without affecting column flow
- Septum purge builtin at 3 ml/min
- Accepts columns unto 1.2-mm od
- Multi mode design includes split and splitless injection
- Splitless purge time variable in 0.01-min elements
Flame Ionization Detector
- Range to 450°C
- Grounded jet and current limited design for operator safety
- Push-button flame ignition
- Sensitive:
- 18 mCoul/gm carbon:nitrogen carrier, 0.018-in. id jet
- 15 mCoul/gm carbon:helium carrier, 0.018-in. id jet
- 22 mCoul/gm carbon:nitrogen carrier, 0.018-in. id capillary jet
- 18 mCoul/gm carbon:helium carrier, 0.018-in. id capillary jet
- Minimum detectable:5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic range:+/- 10% over a 10 7 range with 0.018-in. id jet
- Conditions: column flow 50 ml/min, 45 ml/min H2, 650 ml/min air, propane sample
Flame Photometric Detector
- Range to 300°C
- 100-530 u id capillary columns fit through the transfer line to the base of the flame
- Two combustion modes: air/hydrogen or oxygen/hydrogen
- Detector signal display to view when flame is lit
- Push-button flame ignition
- Sulphur mode: Sensitivity to 20 pg S/second, Range 103, Selectivity 105 g S/g C
- o Phosphorus mode:Sensitivity to 0.9 pg P/second, Range 104, Selectivity 106 g P/g C
Column Oven
- Operating range: 4°C above ambient to 450°C
- Set point entry: 1°C for temperatures, 0.1°C for program rates
- Programming: Rates 0.1 to 70 °C per min, 650 min maximum run time, Three ramps with initial/final holds
This refurbished HP 5890 II GC with Single FID, Dual Split/splitless Inlets is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
Installation and extended warranty available upon request.
Multiple available